Page 22
Find us at www.keysight.com
848xD Series Diode and 8483A Thermocouple Power Sensor Specifications
Calibration factor uncertainties
These thermocouple and diode power sensors provide extraordinary accuracy, stability, and SWR over a wide range of frequencies (100
kHz to 110 GHz) and power levels (–70 to +20 dBm)�
The 8480 Series sensors in the table do not include discontinued models�
Table 13. Typical root sum of squares (rss) uncertainty on the calibration factor data printed on the power sensor.
Frequency (GHz) 8483A 8481D 8485D 8487D R8486D Q8486D
0.0001 1.3 – – – – –
0.0003 1.2 – – – – –
0.001 1.1 – – – – –
0.003 1.2 – – – – –
0.01 1.2 – – – – –
0.03 1.2 – – – – –
0.05 1.2 – – – – –
0.1 1.2 – – – – –
0.3 1.2 – – – – –
1 1.2 0.8 1.4 1.3 – –
2 1.2 0.8 1.4 1.3 – –
4 – 0.8 1.7 1.4 – –
6 – 0.9 1.7 1.4 – –
8 – 1.0 1.7 1.4 – –
10 – 1.1 1.9 1.5 – –
12 – 1.2 1.9 1.5 – –
14 – 1.1 2.0 1.6 – –
16 – 1.5 2.1 1.7 – –
18 – 1.7 2.2 1.7 – –
22 – – 2.7 1.9 – –
26.5 – – 2.8 2.2 3.0 –
28 – – 2.9
1
2.3 3.2 –
30 – – 3.2
1
2.4 3.0 –
33 – – 3.3
1
2.6 3.0 4.2
34.5 – – – 2.6 3.0 4.2
37 – – – 2.7 3.0 4.2
40 – – – 3.0 – 4.2
42 – – – 3.2 – 4.9
44 – – – 2.5 – 5.1
46 – – – 3.8 – 5.5
48 – – – 3.8 – 5.8
50 – – – 5.0 – 6.2
1� These uncertainties only apply to Option 033�