5 Language Reference
66 Series N5700 User’s Guide
SCPI Command Summary
NOTE
Some [optional] commands have been included for clarity. All settings commands
have a corresponding query.
Subsystem Commands
SCPI Command
Description
ABORt
Aborts the triggered action
CALibrate
:CURRent[:LEVel]
Calibrates the output current programming
:DATA <NRf>
Enters the calibration value
:DATE <”SPD”>
Sets the calibration date
:LEVel P1 | P2
Advances to the next calibration step
:PASSword <NRf>
Sets the numeric calibration password
:STATE <Bool> [,<NRf>]
Enables/disables calibration mode
:VOLTage[:LEVel]
Calibrates the output voltage programming
INITiate
[:IMMediate][:TRANsient]
Initiates the trigger system
:CONTinuous[:TRANsient]
Enables/disables continuous triggers
MEASure
[:SCALar]
:CURRent[:DC]?
Returns the measured output current
:VOLTage[:DC]?
Returns the measured output voltage
OUTPut
[:STATe] <Bool>
Enables/disables the specified output
:PON
:STATe RST | AUTO
Programs the Power-On State
:PROTection
:CLEar
Resets latched protection
[SOURce:]
CURRent
[:LEVel]
[:IMMediate][:AMPLitude] <NRf+>
Sets the output current
:TRIGgered[:AMPLitude] <NRf+>
Sets the triggered output current
:PROTection
:STATe <Bool>
Enables/disables over-current protection
VOLTage
[:LEVel]
[:IMMediate][:AMPLitude] <NRf+>
Sets the output voltage
:TRIGgered[:AMPLitude] <NRf+>
Sets the triggered output voltage
:LIMit
:LOW <NRf+>
Sets the low-voltage limit
:PROTection
[:LEVel] <NRf+>
Sets the over-voltage protection level