5 Troubleshooting
114 Model N6705 Service Guide
Backplane PCA Troubleshooting
Figure 5-4. Backplane PCA Troubleshooting Sheet 1
Begin Backplane Assembly Troubleshooting
Inspect the pins on
the connector
Any missing
pins?
Any bent pins?
Any cracks in
solder?
Inspect soldering
around connector
End Backplane Troubleshooting Procedures
1. Replace Backplane Assembly
1. Try to straighten pin
2. Replace Backplane Assembly
1. Re-solder connector
2. Replace Backplane Assembly