Calibration 3
Model N6705 Service Guide 45
Calibration Set-up
The following figures show the calibration test set-ups. Connect all leads
to the Output terminals as shown in the following figures. Twist the + and
- load leads to reduce noise and inductance.
+
1
+
2
AUX VOLTAGE
MEASUREMENT
+
E.
C.
+
A.
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
B.
Current
shunt
+
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter
Enable 4-wire sense.
Press [Settings], then
select Advanced.
Select 4 wire.
+
D.
Keysight 3458A
configured as
an ammeter
Disable 4-wire
sense. LED
indicator is off.
Electronic load
+
+
F.
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter
Disable 4-wire
sense. LED
indicator is off.
External
DC supply
Current
monitor
+
+
G.
Disable 4-wire
sense. LED
indicator is off.
Keysight 3458A
configured as a
voltmeter