3 Calibration
48 Model N6705 Service Guide
Step 6.
Calculate the shunt current (I=V/R) and enter the data.
Enter the calculated value into the data field
and then select Next. When calibration is
complete, select Close.
From the Calibration menu, select
Date. Enter the date in the format
“yyyy/mm/dd”. Then select Insert
Date. Then select Close.
To enter a calibration date:
CAL:DATE “<date>”,
(@1)
Save and Log out of Calibration Mode
Storing calibration constants overwrites the existing ones in non-volatile
memory. If you are not sure you want to permanently store the new
constants, do not Save the data when you exit the calibration mode. The
calibration will then remain unchanged.
From the Calibration menu, select
Save. Then select Save Calibration
for all Outputs. Select Close.
Go [Back] to the Administrative
tools menu and select
Administrator Login. Select Logout,
then Close
To save calibration data:
CAL:SAVE
To exit calibration mode:
CAL:STAT OFF