119
Measuring Digital Communications Signals
Making Adjacent Channel Power (ACP) Measurements
Figure 9-4 Measuring a Third Adjacent Channel
12Set pass/fail limits for
each offset.
Press Meas Setup,
Offset/Limits, Offset, A,
More, Rel Limit (Car),
55, dB, Offset, B, Rel
Limit (Car), 75, dB,
Offset, C, Rel Limit (Car),
60, dB.
13Turn the limit test on. Press Meas Setup, More,
Limit Test (On).
In Figure 9-5 notice that offsets A and
C have passed, however offset B has
failed. Power levels that fall above our
specified –75 dB for offset B, fail. The
offset bar graph and the associated
power level value are shaded red to
identify a failure. The offset limits are
shown as dashed lines.