EasyManua.ls Logo

Keysight N9030A

Keysight N9030A
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
Improving Frequency Resolution and Accuracy
Using a Frequency Counter to Improve Frequency Resolution and Accuracy
Using a Frequency Counter to Improve Frequency Resolution and
Accuracy
This procedure uses the signal analyzer internal frequency counter to increase the
resolution and accuracy of the frequency readout.
Step Action Notes
1 Select the mode. Press Mode, Spectrum Analyzer.
2 Preset the mode. Press Mode Preset.
3 Enable the internal
reference signal.
Press Input/Output, RF
Calibrator, 50 MHz.
4 Set the center frequency
and span.
a. Press FREQ Channel, Center
Freq, 50, MHz
b. Press SPAN X Scale, Span, 80,
MHz.
5 Turn the frequency
counter on.
Press Marker, More, Marker
Count, Counter (On).
The marker counter remains on
until turned off.
Figure 4-1 Using Marker Counter
6 Turn off the marker
counter.
Press Marker, More, Marker
Count, Count (Off).
Or
Press Marker, Off.