390                                                                                                                     Keysight N9927-90001 User’s Guide
ERTA (Extended Range Transmission Analysis) Mode - Option 209
Start Connection Wizard
—Press Continue.
STEP 5: Measure Jumper Cables
These steps can be used to characterize an Input and Output jumper cable that 
may be connect the DUT to the FieldFox. The loss of the characterized jumper 
cables can then be removed from the DUT measurements. The measurements 
are stored to the internal Cables folder using an auto-generated *.csv filename. 
— The Input cable is attached between the Source/splitter output and the 
DUT input.
— The Output cable is attached between the DUT output and the Receiver 
unit. 
Figure 21-1  Input and Output jumper cables are measured on the Receiver unit
If your DUT input AND output is connected directly to the FieldFox, or you 
choose to ignore the loss of BOTH jumper cables, then press Skip this step
. 
Otherwise, press Measure Jumper Cables
. 
Do the following for both the Input and Output jumper cables:
—Press Skip this step 
to skip the cable measurement.
— Otherwise, connect the Input / Output jumper cable as shown below, then 
press Measure Input (or Output) Cable
. 
For optimum flexibility, use the Scale, then More, then Corrections menu to 
measure, save, and recall cable characterizations. These softkeys are 
duplicated from the SA Field Strength - Antenna/Cable characterization. 
Learn more in “Field Strength Measurements” on page 156.