Pulse Measurements Mode – Option 330
Pulse Measurement Settings
How to set Single or Continuous
—Press Sweep 3
— Then choose one of the following:
— Single
   Automatically sets Continuous OFF and causes the FieldFox 
to query the USB Power Sensor ONCE, then hold for the next Single 
key press. Hold is annotated in the upper left corner of the display.
— Continuous OFF ON
 Queries the USB Power Sensor whenever a valid 
trigger signal is present. This is the typical setting when battery 
power is not critical.
You can also use Run / Hold +/-
 to toggle between Single and Continuous.
Resolution
Resolution is the number of individual acquisitions that occur in the USB 
Power Sensor. This data is read by the FieldFox and plotted across the X-axis 
to form a trace. More data points increases measurement resolution. However, 
more data points requires more time to acquire, query, and plot.
How to set Resolution
—Press Sweep 3
—Then Resolution 
— Then choose from the following:
— Low
 240 data points.
— Medium
 1000 data points.
— High
 Calculated value - approximately 8000 data points.
Triggering
For pulse measurements, a trigger signal refers to the event that causes the 
USB Peak Power Sensor to initiate measurement acquisition. It does NOT refer 
to triggering the FieldFox. 
Trigger settings can be used with ALL pulse measurements unless stated 
otherwise.
A Waiting for Trigger message on the FieldFox screen is posted on behalf of 
the USB Power Sensor when it is waiting for a pulse signal at the RF Input 
(Internal) or a TTL signal at the USB Sensor External Trigger input.
How to make Trigger settings
—Press Sweep 3
—Then Trigger