EasyManua.ls Logo

Klein Tools IR1 - General Specifications IR1; Operating Environment and Conditions; Measurement Capabilities and Features; Physical and Power Specifications

Klein Tools IR1
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2
GENERAL SPECIFICATIONS - IR1
Klein Tools IR1 is an entry-level infrared thermometer with targeting
laser. It offers a wide measurement range, a 10:1 distance–to-spot
ratio, a targeting laser, and user-selectable from Fahrenheit to Celsius
temperature scales.
Environment: Indoor or outdoor
Operating Altitude: 6562 ft. (2000 m)
Relative Humidity: <85% non-condensing
Operating Temp: 32° to 122°F (0° to 50°C)
Storage Temp: -4° to 140°F (-20° to 60°C)
Measurement Range: -4° to 752°F (-20° to 400°C)
Units: Settable to °F or °C
Emissivity: 0.95 fixed
Optical Resolution (Distance-to-spot): 10:1
Dimensions: 6" x 4.1" x 1.7" (152 x 104 x 43 mm)
Weight: 7.4 oz. (210 g) including battery
Battery Type: 1 x 9V battery
Battery Life: (Estimates assume 9V Zinc-Carbon Battery)
5 hours continuous use w/laser and backlight on.
Display: Backlit LCD with white backlight
Display Resolution: 0.1°F (0.1° C)
Laser: FDA and IEC Class 2
Wavelength: 630 – 670 nm
Max Power: 1mW
Beam Divergence: 1.5mrad
Standards: Certified to IEC EN 61326-1:2013, EN 61326-2-3:2013,
IEC EN 60825-1:2014
Complies with 21 CFR PART 1040.10 and 1040.11
except for deviation pursuant to
Laser notice No.50,dated June 24, 2007
Pollution degree: 2
Drop Protection: 6.6 ft. (2m)
Specifications subject to change.
ENGLISH
MEASUREMENT SPECIFICATIONS
Response Time: <500ms
Data hold: Yes
Spectral Response: 8000-14000nm
≥32
°F
(≥0
°C
) ±4
°F
(±2
°C
) or ±2%, whichever is greater
<32
°F
(<0
°C
)
±(4
°F
+ 0.1
°F
per degree below 32
°F
)
±(2.0
°C
+ 0.1
°C
per degree below 0
°C
)
Repeatability: ±1.0% of reading or ±2°F(±1°C)
(whichever is greater).
Temperature Correction Coefcient:
±0.1°F per °F (±0.1°C per °C), or ±0.2%
of reading (whichever is greater) when ambient temperature is above or
below 70 – 77°F (21 – 25°C).
Specifications subject to change.