KNF Laboport N86 KN.18 - User Manual

KNF Laboport N86 KN.18
140 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
KNF121435-121437 04/18
Original-Betriebsanleitung, deutsch Für zukünftige Verwendung aufbewahren!
Betriebsanleitung
Diese Betriebsanleitung lesen und beachten!
Laboport
®
Mini-Laborpumpen
N86 KN.18
N86 KT.18
N811 KN.18
N811 KT.18
KNF Neuberger SAS
4, Bld. d’Alsace Z.I.
F-68128 Village-Neuf
Tel. 0033 (0)389 70 35 00
Fax 0033 (0)389 69 92 52
E-mail: info@knf.fr
www.knf.fr
Inhalt Seite
Zu diesem Dokument ....................................................................... 3
2. Verwendung .............................................................................. 4
3. Sicherheit .................................................................................. 5
4. Technische Daten ..................................................................... 7
5. Aufbau und Funktion ............................................................... 12
6. Aufstellen und Anschließen ..................................................... 14
7. Betrieb ..................................................................................... 15
8. Instandhaltung ......................................................................... 17
9. Störungen beheben ................................................................. 23
10. Ersatzteile und Zubehör .......................................................... 25
11. Rücksendungen ...................................................................... 26
12. Unbedenklichkeits- und Dekontaminationserklärung .............. 27
DE
EN
FR
ES
IT

Questions and Answers

  • R
    Ryan MillerAug 11, 2025
    What to do if KNF Water Pump produces no flow?
    • M
      mkiddAug 11, 2025
      If your KNF Water Pump isn't producing any flow, here's what you can check: * Verify that there is voltage in the power source by checking the room fuse and switching it on if necessary. * If the thermal switch has operated due to overheating, disconnect the pump from the mains, allow it to cool, and identify and eliminate the cause of overheating. * Inspect connections and lines for blockages and remove them. * Check external valves and filters. * If condensate has collected in the pump head, detach the condensate source from the pump and flush the pump. * If the diaphragm or valve plates/sealings are worn, replace them.
  • D
    dalton72Aug 27, 2025
    What to do if KNF Water Pump is switched on but does not run and the on/off-switch on the pump is not lit?
    • O
      ohodgeAug 27, 2025
      If your KNF Water Pump is switched on but doesn't run, and the on/off-switch is not lit, check the following: * Ensure the pump is connected to the mains power. * Check the room fuse and switch it on if necessary to ensure there is voltage in the power source. * Remove the pump’s mains plug from the socket, loosen the marked lid on the underside of the pump, and replace the fuse if it is defective.
  • D
    Drew IngramSep 3, 2025
    What to do if KNF Water Pump is switched on, but does not run, the on/off-switch on the pump is lit?
    • T
      Tara BrownSep 3, 2025
      If your KNF Water Pump is switched on but does not run, and the on/off-switch on the pump is lit, it may be due to the thermal switch opening because of overheating. Remove the pump's mains plug from the socket, allow the pump to cool down, and identify and eliminate the cause of the overheating.
  • A
    allen29Nov 18, 2025
    Why is the flow rate low in my KNF Water Pump?
    • S
      Sergio PooleNov 18, 2025
      If the flow rate, pressure, or vacuum is too low in your KNF Water Pump, consider the following: * If condensate has collected in the pump head, detach the condensate source from the pump and flush the pump. * Change the pressure conditions if there is gauge pressure on the pressure side and vacuum or pressure above atmospheric pressure on the suction side. * If pneumatic lines or connection parts have an insufficient cross section, disconnect the pump from the system to determine output values, eliminate throttling if necessary, and use lines or connection parts with a larger cross section if needed. * Check that tubes sit correctly on hose nozzles, replace leaky tubes, and eliminate leaks if leaks occur on connections, lines, or the pump head. * Check connections and lines and...

KNF Laboport N86 KN.18 Specifications

General IconGeneral
Pump TypeDiaphragm Pump
Max Flow Rate18 l/min
Power Consumption90 W
Material (Valves)FFPM
Voltage230V
Frequency50/60 Hz
Gas Temperature5 to 40 °C
Material (Diaphragm)PTFE
MotorAC Motor
Hose Connection10 mm
MaterialPTFE
Ultimate Vacuum100 mbar abs.