KV1000 Constant Temperature Kinematic Viscosity Bath
Operation and Instruction Manual
- 4 -
3.2. Unpacking
Carefully unpack and place the instrument and accessories in a secure location. Use extra care while unpacking the Pyrex® glass
jar. Ensure that all parts listed on the packing list are present. Inspect the unit and all accessories for damage. If any damage is
found, keep all packing materials and immediately report the damage to the carrier. We will assist you with your claim, if requested.
When submitting a claim for shipping damage, request that the carrier inspect the shipping container and equipment. Do not return
goods to Koehler without written authorization.
3.3. Setup
Equipment Placement. Place the instrument on a firm, level table in an area with adequate ventilation or in a hood. Please note
that Koehler does not supply a level with this equipment.
Ventilation. A fume hood or exhaust system is required when operating the unit. Flammable vapors and/or steam are generated
during operation and must not be permitted to accumulate. A canopy-style hood may be used if the height from the top of the unit
to the canopy is 5 feet or less. The exhaust blower should have a rating of 1000 C.F.M or greater.
Power. Connect the line cords to properly fused and grounded receptacles with the correct voltage as indicated in section 1.3 or
on the back of the unit.
WARNING: For safety, disconnect the power when performing any maintenance and/or cleaning. Do NOT turn the power on
unless the bath is filled with the proper medium; otherwise, damage may occur to the unit and the warranty will be void.