Additional calibration points can be entered by clicking Show more. Up to 15 calibration points can be
recorded. Adding calibration points increases the accuracy of determining the wavelengths of spectral lines.
The accuracy of the determination of the calibration curve can be checked by displaying it by selecting
Calibration curve on the Spectrometer calibration panel.
The calibration points must be evenly distributed on each side of the curve with minimal deviation. An
inappropriately chosen calibration point will manifest itself in a significantly larger deviation from the curve. In
this case, it is necessary to change its coordinates or remove the point from the calibration curve.
Advanced setting
In the Advanced settings menu of the calibration panel, it is possible to create a corrected spectral
dependence of the spectrometer's response to the white light source.