5
Over temperature protection Detects overheating of the internal resistor
(used for loop and RCD tests) and of the
current control MOS FET (used for RCD
tests) displaying a warning symbol (
) and
automatically halting further measurements.
Phase angle selector The test can be selected from either the
positive (0
°
) or from the negative (180
°
) half-
cycle of voltage. This will prevent tripping of
some polarised RCDs when loop testing and
may give a more accurate reading when
testing RCDs.
Auto data hold Holds the displayed reading for a time after
the test is complete.
Auto power off Automatically switches the instrument off after
a period of approximately 1 minutes.
V-NE Monitoring Circuit Automatically aborts measurement when the
N-E voltage rises to 50V or greater on RCD
ranges.
Optional Accessory OMA DIEC distribution board or lighting circuit
test lead for LOOP
RCD testing.