What to do if Cold Trap lid is not seated on Labconco CentriVap 7812010?
- BBrian HendersonSep 9, 2025
Hold the lid down until vacuum is initiated if the Cold Trap lid is not seated on your Labconco Laboratory Equipment.
What to do if Cold Trap lid is not seated on Labconco CentriVap 7812010?
Hold the lid down until vacuum is initiated if the Cold Trap lid is not seated on your Labconco Laboratory Equipment.
How to remove ice formed on Cold Trap lid sealing surface of Labconco Laboratory Equipment?
If ice has formed on the Cold Trap lid sealing surface of your Labconco Laboratory Equipment, defrost it and wipe it dry.
What to do if Secondary Trap Cannister is not fitted properly on Labconco Laboratory Equipment?
Tighten all connections to and from the Secondary Trap Cannister on your Labconco Laboratory Equipment if it is not fitted properly.
How to remove moisture from new Secondary Trap in Labconco Laboratory Equipment?
Run the vacuum pump for 24 hours to remove moisture from a new Secondary Trap in your Labconco Laboratory Equipment.
Brand | Labconco |
---|---|
Model | CentriVap 7812010 |
Category | Laboratory Equipment |
Language | English |
Explains the meaning of safety alert symbols and hazard warning words used in the manual.
Details electrical, location, exhaust, and space requirements for installing the CentriVap.
Covers setting up the unit, emergency access, pump installation, exhaust, and electrical connections.
Explains heat boost, control panel functions, and step-by-step operating procedures.
Specific safety measures for operating the CentriVap with hazardous, toxic, or radioactive materials.
Lists common error codes, their causes, and corrective actions for internal system checks.
Provides solutions for common issues like unit not operating, excessive vibration, or poor vacuum.