Chapter 2: Prerequisites
Product Service 1-800-522-7658
5
Vacuum Line Traps
When evaporating acids, it is advisable to include a soda lime acid secondary trap
in the vacuum line. This adds additional protection for the pump.
NOTE: Several components within the CentriVap are made from stainless
steel or aluminum and can be degraded if exposed to acids. Before
evaporating acids, be sure to contact Labconco.
A moisture trap, solvent trap and radioisotope trap are also available. Refer to
Appendix D: CentriVap Accessories for ordering information.
Space Requirements
Refer to Appendix C: CentriVap Specifications for dimensional drawings of the
CentriVap.
No person or any hazardous material should be within 12 inches of the CentriVap
Concentrator while it is operating.