CentriVap® Centrifugal
Concentrators and Cold Traps
Models
78100-00 78110-00 79820-00
78100-01 78110-01 79820-01
78100-02 74600-00 79830-00
78100-03 74600-01 79830-01
79840-00
79840-01
74750-00
74750-01
User’s Manual
LABCONCO CORPORATION
8811 PROSPECT AVENUE
KANSAS CITY, MO 64132-2696
800-821-5525, 816-333-8811
FAX 816-363-0130
E-MAIL labconco@labconco.com
HOME PAGE www.labconco.com
Labconco’s Mascot,
Labby the LABster
Register your product online at
www.labconco.com/productreg.html
or return the attached card and
receive FREE LabbyWear
!

Questions and Answers

  • P
    Pamela SantosAug 12, 2025
    What to do if my Labconco Laboratory Equipment unit will not operate?
    • L
      lawrencecassandraAug 12, 2025
      If your Labconco Laboratory Equipment unit isn't operating, there are several potential causes: * The unit might not be connected to a power receptacle. Ensure it's properly connected. * A circuit breaker may have blown. Correct the electrical issue and reset the circuit breaker by pressing the button. * The lid might be open. Make sure to close the lid securely.
  • C
    Casey GrayAug 20, 2025
    What to do if there is no vacuum or poor vacuum (cont.) in my Labconco Laboratory Equipment?
    • J
      Jennifer RobinsonAug 20, 2025
      If you are experiencing issues such as no vacuum or poor vacuum in your Labconco Laboratory Equipment, consider the following: * Ice may have formed on the Cold Trap lid sealing surface. Defrost it and wipe it dry. * The Secondary Trap Cannister might not be fitted properly. Tighten all connections to and from the Secondary Trap Cannister. * The Secondary Trap Insert could be spent. Replace it with a new insert. * A new Secondary Trap may have moisture in it. Run the vacuum pump for 24 hours to remove the moisture.
  • M
    Mark VelazquezAug 23, 2025
    Why does my Labconco Laboratory Equipment start and shut off?
    • J
      Joseph TorresAug 25, 2025
      If your Labconco Laboratory Equipment starts and then shuts off, it could be due to an improperly calibrated latch optical sensor. To resolve this, unplug the power cord, wait 10 seconds, and then plug the power cord back in.
  • S
    Sarah KnightAug 27, 2025
    Why is frequent oil change needed in pump of my Labconco Laboratory Equipment?
    • A
      Aaron SnyderAug 27, 2025
      If you find that the pump in your Labconco Laboratory Equipment requires frequent oil changes, it could be due to a few reasons: * The Secondary Trap insert may be spent. Change the insert often. * The Cold Trap might not be emptied regularly. Empty the traps (glass or stainless steel) after each run and replace them. * The vacuum might be too strong for a chemical diaphragm pump. Use a Secondary Trap insert and chemical diaphragm pump.
  • C
    campbellteresaAug 31, 2025
    Why is the recovery of condensate in Cold Trap less than normally expected in my Labconco Laboratory Equipment?
    • A
      April RiceAug 31, 2025
      If the recovery of condensate in the Cold Trap of your Labconco Laboratory Equipment is less than expected, consider the following: * The Cold Trap may not be turned ON. Check to ensure the switch is ON and the condensing unit fan is moving air out of the rear of the cabinet. * The Cold Trap may not be cooling down. Turn the vacuum pump OFF and allow the Cold Trap to cool for at least 30 minutes to reach temperature.
  • S
    Sarah RiceSep 4, 2025
    What causes excessive vibration in Labconco Laboratory Equipment?
    • C
      Cindy KingSep 4, 2025
      Excessive vibration in Labconco Laboratory Equipment is caused by sample tubes not located symmetrically in rotor. To solve this, reposition sample tubes.
  • B
    brittney73Sep 8, 2025
    Why is there sample odor in lab when using Labconco Laboratory Equipment?
    • M
      Michael HarrisSep 8, 2025
      The presence of sample odor in the lab when using Labconco Laboratory Equipment is caused by the vent hose exhausting into the lab area. To resolve this, redirect the hose to a fume hood.
  • J
    Jason SwansonNov 15, 2025
    Why is there no vacuum in my Labconco Laboratory Equipment?
    • G
      gardnerjenniferNov 15, 2025
      If you're experiencing a lack of vacuum in your Labconco Laboratory Equipment, consider these possible causes and solutions: * The pump might not be turned on. Ensure the pump is switched on. * The control valve could be open. Check the control valve and adjust as necessary. * There may be leaks in the lines, connectors, or gasket. Locate and repair any leaks. * Foreign material might be present on the lid gasket. Clean the gasket and lid thoroughly. * The pump might not be functioning properly. Check the pump by positioning the vacuum gauge closer to it and closing off the rest of the system. Inspect the pump oil for cloudiness or particles and change it. If the pump is faulty, seek authorized service or replace it. * The Cold Trap or Concentrator gasket may not be sealing co...

Summary

Chapter 2: Prerequisites

Electrical Requirements

Details necessary electrical supply and circuit breaker specifications.

Location and Exhaust Requirements

Specifies ideal placement and ventilation for safe operation.

Vacuum Pump Requirements

Outlines requirements for compatible vacuum pumps and specifications.

Chapter 3: Getting Started

Electrical Connection

Guides on connecting power cords for the concentrator, cold trap, and pump.

Solvent Safety Precautions

Crucial safety guidelines for handling solvents with the unit.

Chapter 4: Using Your CentriVap

CentriVap Controls Overview

Explains the function of each control on the CentriVap panel.

Operating The CentriVap

Step-by-step guide on using the control panel for operation and running cycles.

Chapter 6: Modifying Your CentriVap

Appendix C: CentriVap Specifications

Electrical Specifications

Details amperage, frequency, and phase for electrical requirements.

Wiring Diagram Concentrator (115 V)

Electrical wiring schematic for the 115V Concentrator.

Wiring Diagram Concentrator (230 V)

Electrical wiring schematic for the 230V Concentrator.

Wiring Diagram Cold Trap (115 V;230 V)

Wiring schematic for both 115V and 230V Cold Trap models.

Evaporation Rates

Tables showing evaporation rates for various solvents and conditions.

Labconco CentriVap 78110-00 Specifications

General IconGeneral
Model78110-00
TypeCentrifugal Concentrator
BrandLabconco
Vacuum PumpNot included
Rotary EvaporatorNo
ApplicationConcentration of samples