Protector
®
XStream
®
Laboratory Fume Hoods
User’s Manual
LABCONCO CORPORATION
8811 PROSPECT AVENUE
KANSAS CITY, MO 64132-2696
800-821-5525, 816-333-8811
FAX 816-363-0130
E-MAIL labconco@labconco.com
HOME PAGE www.labconco.com
Labconco’s Mascot,
Labby the LABster
Register your product online at
www.labconco.com/productreg.html
or return the attached card and
receive FREE LabbyWear
!
Models
9840400, 9840401, 9840402, 9840403,
9840500, 9840501, 9840502, 9840503,
9840600, 9840601, 9840602, 9840603,
9840800, 9840801, 9840802, 9840803

Table of Contents

Questions and Answers

  • J
    Julia PhelpsAug 7, 2025
    What to do if there are contaminants outside of my Labconco Laboratory Equipment fume hood?
    • A
      Antonio WilkinsonAug 7, 2025
      If you're experiencing contaminants outside of your fume hood, several factors could be at play. First, check for any obstructions in the baffle air slots or the exhaust outlet; remove baffles to ensure all air slots and the exhaust outlet are unobstructed. Also, external factors might be disrupting the fume or acting as a source of contamination. Finally, if you have an XStream Fume hood, it may have an improper face velocity and should be re-certified, and the remote blower exhaust system should be checked. The XStream Hood should have an average face velocity of 60-100 fpm.
  • D
    Dakota JohnsonAug 12, 2025
    Why do the electrical duplex outlets on my Labconco Laboratory Equipment no longer have power?
    • M
      Mrs. Lauren RamosAug 12, 2025
      If the electrical duplex outlets on your equipment have no power, it could be due to a couple of reasons. Either the wires are not connected or the duplex is faulty, so check the wire connections or replace the duplex. Alternatively, the circuit breakers in the building's electrical supply may have tripped; in this case, reset the circuit breakers.
  • J
    Jonathon PetersonAug 19, 2025
    How to fix a Labconco Protector XStream 9840400 vertical sash that no longer operates smoothly?
    • J
      Jason BrooksAug 19, 2025
      If your vertical sash isn't operating smoothly, several issues could be the cause. Inspect the cable for fraying or damage to the plastic protection, and replace it immediately if worn or damaged to prevent injury. Also, the pulley bearing may be damaged, requiring replacement or greasing. Check if the cable has slipped off the pulleys and re-install it, replacing it immediately if damaged. Additionally, weight pulleys might be broken and need replacement, or horizontal glass panels may have come off the tracks, requiring re-installation.
  • D
    Daniel SandersAug 25, 2025
    What to do if service valves on my Labconco Protector XStream 9840400 show no PSI?
    • L
      levinemartinAug 25, 2025
      If your service valves show no PSI, there are several things to investigate. Start by inspecting the building supply shut off valves for any issues. If the valve no longer operates, replace it and check for leaks. Finally, inspect the supply or outlet lines for leaks and fix any leaking plumbing connections.
  • M
    Michael RandallAug 30, 2025
    How to fix lights that won't operate on my Labconco Protector XStream 9840400 Laboratory Equipment, even though the fume hood blower works?
    • A
      Angel LynchAug 30, 2025
      If your fume hood blower is working but the lights are not, it could be due to a couple of reasons. First, inspect the lamp wiring to ensure it is properly connected. If the wiring is intact, the lamp ballasts may be defective and need to be replaced.

Summary

Copyright Information

Warranty

Details Labconco's warranty policy for parts, factory workmanship, and product use.

Limitation of Liability

User's responsibility for regulations and Labconco's liability limits.

Contacting Labconco Corporation

Provides contact details for customer and product service assistance.

CHAPTER 1 INTRODUCTION

About This Manual

Explains the manual's purpose, structure, and conventions for user understanding.

Typographical Conventions

Defines icons, formatting, and symbols used throughout the manual for clarity.

Your Next Step

Guides users on how to navigate the manual based on their specific needs.

CHAPTER 2 PREREQUISITES

Location Requirements

Specifies ideal placement of the fume hood away from disruptions.

Support Requirements

Details the need for permanent, stationary support structures for the hood.

Exhaust Requirements

Outlines duct connection size and transport velocities for exhaust systems.

Electrical Requirements

Explains internal wiring for light assembly and connection to a single point junction box.

Service Line Requirements

Specifies requirements for service lines, including size, material, and shut-off valves.

Space Requirements

Refers to Appendix B for detailed dimensions of different fume hood models.

CHAPTER 3 GETTING STARTED

Unpacking Your Laboratory Fume Hood

Instructions for removing packaging, inspecting for damage, and handling returns.

Removing the Shipping Skid

Details how to safely remove the fume hood from its shipping skid.

Sash Weight Release

Explains how to secure and release the sash weight for safe transport.

Install the Protector XStream Hood on a Supporting Structure and Work Surface

Guides on safely installing the hood onto its support and work surface.

Connecting to the Protector XStream Hood Exhaust System

Instructions for connecting the hood to the exhaust ductwork system.

Connecting the Electrical Supply Source to the Protector XStream Fume Hood

Details how to connect the hood to the building's electrical supply.

Connecting the Service Lines to the Protector XStream Fume Hood

Explains how to connect gas, water, or vacuum lines to the fume hood fixtures.

Sealing the Protector XStream Hood to the Work Surface

Instructions for sealing the hood to the work surface to prevent spills from collecting.

Certifying the Protector XStream Fume Hood

Outlines procedures for certifying the fume hood's performance and airflow.

CHAPTER 4 HIGH PERFORMANCE FEATURES AND SAFETY PRECAUTIONS

High Performance Features

Describes unique design elements that enhance containment, safety, and energy savings.

Safety Precautions

Provides essential safety guidelines for operating and maintaining the fume hood.

CHAPTER 5 USING YOUR PROTECTOR XSTREAM FUME HOOD

Operating the Vertical-Rising Sash

Instructions for operating the hood's vertical sash mechanism.

Operating the A-Style Combination Sash

Guidance on using the optional A-Style sash for energy conservation.

Operating the Blower

Instructions on how to activate and check the fume hood blower.

Operating the Lights

How to turn on and use the interior fluorescent lighting.

Working in your Protector XStream Fume Hood

Covers planning, start-up, material loading, work techniques, purging, and shutdown procedures.

CHAPTER 6 MAINTAINING YOUR PROTECTOR XSTREAM FUME HOOD

Routine Maintenance Schedule

Outlines weekly, monthly, and annual maintenance tasks for the fume hood.

Front Panel Removal

Steps to remove the front panel for access to components.

Changing the Fluorescent Lamp

Detailed instructions for replacing the fume hood's fluorescent lamp.

CHAPTER 7 MODIFYING YOUR PROTECTOR XSTREAM FUME HOOD

Installing Work Surfaces

Information on the required work surface for proper hood performance.

Installing Ceiling Enclosures above the Fume Hood

Instructions for mounting ceiling enclosures to close the gap between hood and ceiling.

Installing Rear Panels Behind the Fume Hood

Guidance on adding rear panels for island installations.

Installing Additional Service Fixtures

How to install extra service fixtures in sidewalls and corner posts.

Installing Guardian Digital Airflow Monitor or Guardian Jr. Airflow Monitor

Instructions for installing airflow monitoring accessories.

Distillation Grids - Field Installation

Details on strategically placing distillation grids for specific apparatus mounting.

Installing an Electrical Duplex Outlet

Procedures for installing field-mounted electrical duplex outlets.

CHAPTER 8 TROUBLESHOOTING

Remote blower and lights wont operate

Troubleshooting steps for issues with the blower and lights not working.

Remote blower wont operate, but lights work

Solutions for when the blower is off, but lights are functional.

Fume hood blower operates but lights will not operate

Troubleshooting steps when the blower works, but the lights do not.

Contaminants outside of fume hood

Addresses issues causing contaminants to escape the fume hood.

Vertical sash no longer operates smoothly

Solutions for when the vertical sash movement becomes difficult.

Combination A-Style Sash no longer operates smoothly

Troubleshooting steps for smooth operation issues with the A-Style sash.

Electrical duplex outlets no longer have power

Steps to resolve power issues with electrical duplex outlets.

Service valves no longer operate

Troubleshooting guide for non-operational service valves.

APPENDIX A PROTECTOR XSTREAM COMPONENTS

Protector XStream Replacement Parts

Lists part numbers and descriptions for various service components.

APPENDIX B PROTECTOR XSTREAM DIMENSIONS

XStream Dimensional Drawing

Provides detailed diagrams and measurements for different fume hood models.

APPENDIX C PROTECTOR XSTREAM FUME HOOD SPECIFICATIONS

Environmental Conditions

Details acceptable environmental parameters for operating the fume hood.

APPENDIX D REFERENCES

Laboratory Ventilation Standards

Lists relevant regulatory standards and recommendations for lab ventilation.

Occupational Health and Safety

Provides contact information for the U.S. Department of Labor's safety division.

Industrial Ventilation-ACGIH

Lists ACGIH guidelines for fume hood face velocities and ventilation practices.

ASHRAE 110-1995 Method of Testing Performance of Fume Hoods

Describes ASHRAE standards for testing fume hood containment characteristics.

ANSI Z9.5-1993 Laboratory Standard

Covers laboratory ventilation systems, including discharge and monitoring requirements.

SEFA 1-2002

SEFA standards for fume hood face velocities based on chemical toxicity.

NFPA 45 - 2002 Fire Protection for Laboratories Using Chemicals

Fire protection guidelines for laboratories, including fume hood exhaust and construction.

NFPA 30 - 2000 Flammable and Combustible Liquids Code

Regulations concerning flammable and combustible liquids, including cabinet requirements.

National Fire Protection Association

Contact information for the National Fire Protection Association.

General References

Lists various books and publications related to laboratory design and safety.

DECLARATION OF CONFORMITY

Labconco Protector XStream 9840400 Specifications

General IconGeneral
TypeFume Hood
ModelProtector XStream 9840400
Airflow100 fpm
Airflow at 100 fpm100 fpm
Width (exterior)48 in
Electrical120V, 60Hz
ConstructionEpoxy-coated steel
SashVertical
LightingLED