16
Product Service 1-800-522-765
8
Chapter 3: Getting Started
through the work surface to make these connections to the service
fixtures.
NOTE: Inspect all fittings for leakage. Tighten
the fittings slightly if needed.
Should access to the hood plumbing fixture bodies be required,
remove the service access plate on the hood front corner posts by
loosening their individual screws. (See item 11 Figure A-1, page
50 and 51) The valve body will now be fully exposed for any
service work that may be necessary. The service fixtures supplied
on your laboratory hood are designed for use with the following
services:
Air Hot Water Vacuum
Cold Water Natural Gas – See Caution Below
Sealing the Protector XStream
Hood to the Work Surface
When the hood has been set in place, ducted, wired, and plumbed,
it should be sealed at the work surface to prevent spilled materials
from collecting under the walls of the hood. Materials such as
silicone sealants are recommended to seal the hood structure.
!
CAUTION: Do not use oxygen with any standard
service fixture. Contact Labconco Customer
Service for oxygen fixture information.
!
WARNING: Contact Labconco Customer
Service directly before using any service other
than those listed above in these valves to assure
full compatibility.
!
CAUTION: Natural gas should be used only in
the service fixture that has been pre-plumbed
with brass tubing. Sulfur content of the gas
could cause deterioration of standard copper
supply lines.
!

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector XStream 9840400 and is the answer not in the manual?

Labconco Protector XStream 9840400 Specifications

General IconGeneral
BrandLabconco
ModelProtector XStream 9840400
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals