Chapter 6: Modifying Your CentriVap
Installing a Secondary
Chemical Trap
An accessory secondary chemical trap is available to
minimize the exhausting of solvents into the
atmosphere. It may be attached to either side of the
Cold Trap. After selecting the side, remove the two
small plastic hole plugs. Attach the bracket to the side
of the Cold Trap housing using the screws provided.
Attach the hose from the Cold Trap Lid Assembly to
the “out” connector of the canister housing. Connect
another hose from the remaining fitting on the
Secondary Trap to the inlet port on the vacuum pump.
Clamp hoses securely. Unscrew the clear bowl of the
canister housing from the head. Remove both the upper
and lower caps from the filter cartridge and insert the
small end of the cartridge into the hole in the center of
the head. Reinstall the clear bowl.
Clear Cannister
Hose
Bracket