52
Product Service Domestic 1-800-522-7658, International 816-333-8811
Appendix C: CentriVap Specifications
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to
40°C).
Maximum relative humidity: 80% for temperatures
up to 88°F (31°C), decreasing linearly to 50%
relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed
±10% of the nominal voltage.
Transient overvoltages according to Installation
Categories II (Overvoltage Categories per IEC
1010). Temporary voltage spikes on the AC input
line that may be as high as 1500V for 115V models
and 2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e.,
where normally only non-conductive atmospheres
are present). Occasionally, however, a temporary
conductivity caused by condensation must be
expected, in accordance with IEC 664.