25
Product Service: Domestic 1-800-522-7658, International 816-333-8811
Chapter 4: Using Your Freeze Drye
r
6. Ampules may be flame sealed while connected
to a valve by using a sealing torch. Care must
be taken not to burn the valve. An insulation
material placed between the valve and the torch
is recommended.
Alarms and Informational
Messages
If any of the following events occur during a run,
the alarm LED will illuminate to notify the user.
1. Power Interruption – If the power to control
system is interrupted either by a power failure
or by turning the main power switch off during
a freeze dry process, the alarm will be set.
When power is restored the freeze dryer will
continue to run and the alarm indicator will
flash. The display will continue to show the
operating parameters of the system. Press
MENU to display the alarm “POWER FAIL.”
Pressing MENU multiple times will identify
N
o
t
e location of taper on
knob in vent position.