Chapter 4: Lyo-Works Operating System
Product Service 816-333-8811 or 1-800-522-7658
22
Original instructions
Shelf Temp. - May be set anywhere from +50°C to 55°C in 1°C
increments. There is also a Max Cold (MC) setting that allows the system
to reach its lowest possible temperature (approximately -75°C). Max Cold
should only be used for pre-freezing samples, not for freeze drying, since
the collector coil is not being cooled in this mode.
Time This time represents how long the system will hold at the Shelf
Temp. that has been programmed for each Step. The time can be set from
00:01 to 99:59 (1 min to 99 hrs 59 min). It can also be set to an indefinite
amount of time, which is represented by the infinity symbol (∞).
Vacuum Vacuum control can be set from 0.000 to 1.500 mbar.
Illustration of a simple 3 Step Program
Data Logging Screen
If no data log file has been selected the “New” button will be the only active
button along the bottom of the screen. A data log file can be selected from the list
by pressing the data log file name. When a data log file has been selected, the row

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco FreeZone Triad 794001010 and is the answer not in the manual?

Labconco FreeZone Triad 794001010 Specifications

General IconGeneral
BrandLabconco
ModelFreeZone Triad 794001010
CategoryDryer
LanguageEnglish

Related product manuals