Chapter 3: Getting Started
Product Service 1-800-522-7658
12
To install the filters, push the stacked filters to the stop in the rear of each
filter compartment. After loading the stacked filters into the compartments,
close the gas spring cylinders by pulling the handle down and compressing
the filter gaskets. Filter installation is complete.
Figure 3-2
Filter Installation
Connect the Electrical Supply
Simply connect the 115V power cord supplied to the IEC electrical supply
plug on the back of the enclosure. For 230V, the same procedure applies
except it is shipped without a plug. Install the appropriate plug for your
electrical specifications per local codes. (See Appendix C for amperage.)
Set Filter Life and Filter Check Timers
Once the filters are installed and electrical power is connected, it is necessary
to set the Filter Life Timers in the Filter Setup menus described in Chapter 5:
Using the Ductless Enclosure. Consult the separate Paramount Chemical
Guide, Appendix E: Estimated Filter Life Calculation, and/or Labconco
Specialist to determine the filter capacity. Through the use of all three
sources, set the Filter Life Timers.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Paramount 69632 Series and is the answer not in the manual?

Labconco Paramount 69632 Series Specifications

General IconGeneral
BrandLabconco
ModelParamount 69632 Series
CategoryEnclosure
LanguageEnglish

Related product manuals