n
The following examples are useful for estimating filter life so the Filter Life
Timers can be set properly.
Organic Vapor Carbon Filter Life
Consult the separate Paramount Chemical Guide and/or Labconco Specialist
to determine the filter capacity for your chemical. For example, assume you
are working with 10 ppm acetone in a 3' Paramount loaded with the required
four filters. Acetone has a specific gravity SPGR = 0.79 with a molecular
weight MW = 58.1. A 3' Paramount at 80 fpm uses 175 CFM or 4956 LPM.
The evaporation rate in ml/min can be solved by the following equation:
PPM = (SPGR/MW) * (24400000/Airflow (LPM)) * Evap. Rate (ml/min).
For 10 ppm acetone or 90 ml lost per week during 2 hours use per day, the
equations are solved below:
10 = (0.79/58.1) * (24400000/4956) * Evap. Rate (ml/min)
Evap. Rate = 0.15 ml/min = 90 ml/week * 1 week /10 hours use *
10 hours/600 minutes
Evap. Rate in (g/min) = 0.15 ml/min * 0.79 g/ml = 0.12 g/min
There are 30 pounds of activated carbon in a 3' Paramount which equals
13,500 g. Thus, the carbon Filter Life Time is calculated for acetone with a
2% filter capacity at (13,500 g/0.12 g/min) / (1 hour/60 min) * 2% capacity =
38 hours. The Final Life Timer should be set for 40 hours, rounded up from
38 hours. The Check Life Timer should be set for 20-25% intervals or every
10 hours and the carbon filters can be validated every 10 hours to verify safe
filter use.
Other chemicals such as ethyl acetate at 9% filter capacity and toluene at
20% filter capacity have much longer filter lives. For example, ethyl acetate
at 10 ppm used in the 3' Paramount will have a Filter Life of 120 hours.
Furthermore, toluene at 10 ppm used in the 3' Paramount will have a Filter
Life of 240 hours. These three chemicals are summarized in the table that
follows to show the Filter Life ranges and you should consult the Labconco
Specialist and/or Paramount Chemical Guide for further information on other
chemicals.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Paramount 69632 Series and is the answer not in the manual?

Labconco Paramount 69632 Series Specifications

General IconGeneral
BrandLabconco
ModelParamount 69632 Series
CategoryEnclosure
LanguageEnglish

Related product manuals