damage the stainless steel.
Heavy duty polyvinyl chloride (PVC) or rubber gloves which provide suitable protection for the hands
and wrists and which can be disinfected for re-use should be worn when cleaning the cabinet.
Procedure:
1. Leave the cabinet running at normal speed. Turn on the light.
2. Remove objects and appliances from the work chamber.
3. Lift up the table tops one by one and clean them carefully on both sides. Remove them from the
working chamber.
4. Remove all loose objects from the trough, flock filter or the prefilter on Mars Pro.
5. Swab the trough carefully on surfaces, along edges and in corners.
6. Clean the back wall and the side windows.
7. Raise the sliding front window to the highest position using the control panel and open the hinged
front cover.
8. Clean the window on the inner side.
9. Close the front cover.
10. Put the clean tabletops back in their place.
11. Let the fan run for 10 minutes.
11.1 Weekly
Wipe the outside of the unit with a mild household cleaner. Anti-static spray may be used to clean
the front window.
11.2 Regularly
Reliable function of the cabinet and compliance with standards are based on the following
conditions:
1. Correct air velocities.
2. Efficiency of the HEPA filters installed.
3. Correctly adjusted alarm limits.
4. Tightness of the construction.
11.3 Decontamination by use of formaldehyde
The purpose of this section is to state guidelines regarding decontamination by using formaldehyde.
Certification is required in certain countries for decontamination by using formaldehyde.
Procedure for decontamination by use of formaldehyde

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaboGene SCANLAF Mars GS Series and is the answer not in the manual?

LaboGene SCANLAF Mars GS Series Specifications

General IconGeneral
BrandLaboGene
ModelSCANLAF Mars GS Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals