SX-01635-000, Rev. 5 26
Appendix A Specifications
Radiometer Performance
Temperature Coefficient: +(0.02%of reading + 0.5 counts) per °C
Input Bias Current: 0.2 pA maximum
Current Input Characteristics at 25°C
Range Resolution Accuracy (% reading + counts)
24 hr period
Accuracy (% reading + counts)
1 yr period
1 nA 0.1 pA 0.05 + 2 0.1 + 5
10 nA 1 pA 0.05 + 1 0.1 + 2
100 nA 10 pA 0.05 + 1 0.1 + 2
1 µA 100 pA 0.05 + 1 0.1 + 2
10 µA 1 nA 0.05 + 1 0.1 + 2
100 µA 10 nA 0.05 + 1 0.1 + 2
1 mA 100 nA 0.05 + 1 0.1 + 2
Voltage Input Characteristics at 25°C
Range Resolution Accuracy (% reading + counts)
24 hr period
Accuracy (% reading + counts)
1 yr period
10 mV 1 µV 0.05 + 2 0.1 + 5
100 mV 10 µ V 0.05 + 1 0.1 + 2
1 V 100 µV 0.05 + 1 0.1 + 2
10 V 1 mV 0.05 + 1 0.1 + 2
50 V 5 mV 0.05 + 1 0.1 + 2
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com