9 Specifications
English
15
Electromagnetic Emissions Tests
Emissions Test Standard or test Compliance
Conducted
and radiated
RF emissions
CISPR 11 Group 1 Class B
Harmonic distortion IEC 61000-3-2 Complies
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3 Complies
Electromagnetic Immunity Tests
Immunity Test Standard or test Compliance Level
Electrostatic
discharge
IEC 61000-4-2 ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Radiated RF EM fields
IEC 61000-4-3 10 V/m
80 MHz- 2.7 GHz
80% AM at 1 kHz
Proximity fields from RF Wireless
communications equipment
IEC 61000-4-3 380-390 MHz: 27V/m
430-470 MHz: 28V/m
704-787 MHz: 9V/m
800-960 MHz: 28V/m
1700-1990 MHz: 28V/m
2400-2470 MHz: 28V/m
5100-5800 MHz: 9V/m
Electrical fast transients / bursts*
IEC 61000-4-4 ± 2 kV
Surges Line-to-line* IEC 61000-4-5 ± 0,5 kV, ± 1 kV
Conducted
disturbances
induced by RF fields*
IEC 61000-4-6 3 V 0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands between
0,15 MHz and 80 MHz
Voltage dips* IEC 61000-4-11 0% UT; 0,5 cycle At 0°, 45°, 90°, 135°, 180°,
225°, 270 ° and 315°
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0°
Voltage
interruptions*
IEC 61000-4-11 0% UT; 250/300 cycle
Rated power
frequency magnetic fields
IEC 61000-4-8 30 A/m
50 Hz or 60 Hz
Electrical transient conduction along
supply lines, DC
power port
ISO 7637-2 Test pulse severity level:
III in Table A2 of
ISO 7637-2
*AC/DC adapter only.
Filter housing K-Resin
300 ml Disposable Canister PC Internal Filter: PE
Patient Port PP
Patient Tube PVC
Wire Stands Steel, PVC