Electromagnetic Conformity
Laerdal Suction Unit is intended for use in the following environments: Professional
Healthcare Facility environment and Emergency Medical Services environment.
Essential performance of the LSU is identified as connection of the patient hos to
the exhaust outlet must be prevented by coding of the outlet and identifying the
exhaust using a label on the device. EMC disturbances cannot affect this behavior.
No particular actions are required to maintain safety and performance with
regard to electromagnetic disturbances for the expected service life.
Warning
Use of this equipment adjacent to or stacked with other equipment should be avoide
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
Warning
Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
Warning
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the LSU, including cables specified by the Laerdal Medical. Otherwise,
degradation of the performance of this equipment could result.