7
door is opened, the sterilizer will stop warming up the chamber and steam generator. The longest time to keep
temperature is 8 hours.
This function can shorten the time of your next sterilization program.
✓ OFF: Disable the function.
4.1.3 ■W/OFF:Indicate the status of function of “Preheat”
The relative menu:
❖ “ADV PREHEAT:ON/OFF” .
✓ ON: It means if user has started a sterilization program, the autoclave will not start the next step until the temperature
of chamber reaches 50℃.
User can set up it in menu: “ADV PREHEAT:ON/OFF”
4.1.4 COUNT:Times of already running sterilization program
i.e. 00023 indicate the sterilizer have ran 23 times
B&D/helix test and vacuum test are not counted.
4.1.5 2 Kpa:
It indicates that the pressure in the chamber is 2 Kpa;
when the autoclave’s door is opened, this pressure is the local air pressure.
4.1.6 14:09:00: Time
User can set it in menu: “ADV DATE/TIME”
4.1.7 26-11-2011: Date
User can set it in menu: “ADV DATE/TIME”
4.1.8 USER: User menu
All programs are in this menu, User can select the program in this menu.
4.1.9 ADV: Advance menu/Set-up Menu
User can change options and set the parameters in this menu
4.1.10 SERV: Serve menu
This menu is for maintenance. It requires password. Only the distributor can use this menu. Please don’t change the
parameters inside this menu without the approval of a professional engineer.
4.1.11 134℃/solid : Current Sterilization Program
It can also record the last sterilization program. Users need not to enter USER menu to select the same program.
If user set ■W/ON, the whole sterilization cycle might be very long. In some countries, it is
required. Please check your local standards before seting up this function.