15
result.
When putting into and taking out the loads, you shall use the tool attached to prevent your hands from touching the
hot surface.
6.3.3 Set up the parameters
Set the parametersKEEP TEMP, PRINTER, PREHEAT if in need.
6.3.4 Close the autoclaves door
If the door is not locked, the LCD will show “Please close the door”.
6.3.5 Choose sterilization program, then start
After the door is closed, choose the relative sterilization program according to the kinds of loads in the chamber.
All the sterilization programs are in “USER” menu. You can press UP or DOWN button to move the cursor to USER menu,
then press OK to enter the menu. Inside the menu, you can press UP or DOWN to choose the sterilization program. After
that, you can press OK. Then, the LCD will show “please push the start key to start” , after seeing that, you can press
“START/STOP” to run the program you selected.
The whole sterilization cycles will be automatically finished. You need not do anything during the programs. The period of
whole cycle is influenced by the initial temperature, the kinds and quantity of the loads.
When the cycle is finished, LCD will show END together with 3 beeps. The user can open the door and take out loads
inside now.
6.3.6 After using, you shall turn off the autoclave. If you will not use it for a long time, you shall disconnect the power
supply.
6.3.7 If The LCD shows “PLEASE DRAIN WATER FROM WASTER WATER TANK!”this mean the waster water tank is
full, you must drain the waste water.
7. Abnormal Situations
The sterilizer will give alarm, release pressure and stop heating automatically if it has any abnormal situations during
If the temperature and pressure inside the chamber are high, the door will be difficult to be closed.
We suggest to close the door quickly, or, you can wait 30 seconds, then close the door.
The user shall fill in water promptly if there is a low-water alert. Otherwise it will show E08 or E9
error alarm.
You shall use a tray-holder tool when taking out loads to prevent scald.
Don’t open the door until the pressure is within “-05~05”.
To ensure the effectiveness of sterilization, during daily use, we suggest putting test paper or
pouches with indicators together with the loads into the sterilization chamber.
The waste water might be very hot, please be careful when you are draining them.