19
8.7 Clean seal ring regularly
The user needs to clean the seal ring regularly by distilled water. If leakage still happens after cleaning, the user may have
to replace the seal ring.
8.8 Replace the seal ring
ToolThe user needs a flat screwdriver.
A. Hold the seal ring by a hand and use another one to hold a screwdriver carefully to separate the door and the seal
ring. Then take the seal ring out slowly.
B. After the user takes the seal ring out, clean and check it. If it is damaged, the user must replace it.
C. After the seal ring is cleaned, assemble it.
D. Attention: if the user finds it hard to put the seal ring back, use screwdriver press it carefully until it is done.
8.9 Regular check the safety valve. If the safety valve is invalid, it must be replaced
Replace the safety valve:
(pic8)
1, Removing the part 1 in pic8, then removing the pipe which connects the safety valve.
2, Removing the screw ( part2 in pic8);
3,Replace the new safety valve.
9. Transportation and Storage
The new safety valve shall be same model. If you need any help, please contact your local
distributor or us.
Only maintain the autoclave when it cools down by well trained person.