3.2 Sterilization Programs
The data have “*” can be adjusted
4. Operation Panel
4.1. LCD screen
4.1.1 ■P/ON:Indicate the status of printer
The relative set-up menu for printer:
❖ ADV PRINTER:ON/OFF
✓ ON : the printer will work
✓ OFF: the printer will not work.
4.1.2 ■K/OFF:Indicate the status of function of “Keep temperature”
The relative menu:
❖ ADV KEEP TEMP:ON/OFF” .
✓ ON : indicate that the sterilizer will warm up the chamber and steam generator to keep the temperature. When the
Process
Vacuum times
Sterilizing time
Sterilization
temperature
Sterilization
pressure
Vacuum persist
time
Dry time
121℃/SOLID
1 *
20min *
121℃
110kPa
3 *
121℃/POROUS
3 *
20min *
121℃
110kPa
7 *
121℃/HOLLOW
3 *
20min *
121℃
110kPa
10 *
134℃/SOLID
1 *
4min *
134℃
210kPa
3 *
134℃/PROOUS
3 *
4min *
134℃
210kPa
7 *
134℃/HOLLOW
3 *
4min *
134℃
210kPa
10 *
USER DEFINED
3 *
5min *
134℃*
210kPa
10 *
B-D TEST
3
4min
134℃
210kPa
7
VACUUM TEST
-80kPa
15min
CLEAN PROCESS
3
5min
105℃
20kPa
10
PRIONEN
3
19min
135℃
210kPa
10
Button
LCD screen
Sterilization
Program
Button
B&D Test: The display shows 4min,but normally it takes 3.5min.