The Steam sterilizer is intended for use in the electromagnetic environment specified below. The customer or the user of
the Steam sterilizer should assure that it is used in such an environment.
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines
2 kV for power
supply lines
1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
1 kV line(s) to line(s)
2 kV line(s) to earth
1 kV line(s) to line(s)
2 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Steam sterilizer requires continued
operation during power
mains interruptions, it is
recommended that the Steam
sterilizer be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.