22
Before doing any precision levelling it is
advised to check the calibration of the
instrument.
Set up the laser on a tripod at about 30m
facing a wall or staff with a detector.
Allow the instrument to level.
Detect and note laser position on the
wall or staff.
Without moving the tripod rotate the
instrument 180°.
Detect and note laser position on the
wall or staff.
Calculate the distance between the two
readings.
The difference should be within 6mm at 30m.
If the instrument is out of calibration it is
advised to send it in to Spot-on.
(see www.spoton.com.au for details)