HML 210
6
TECHNICAL CHARACTERISTICS
ELECTRIC CHARACTERISTICS
ENVIRONMENTAL USE CONDITIONS
Data plate
On the data plate there is the followings technical information:
• MODEL: Converter Model
• S/N: Serial Number of the converter
• PW.TYPE: Not used
• PW.SUPPLY: Power Supply
• MAX.POWER: Maximum power consumption
• Hz: Frequency (AC)
• T: Operating temperature
• IP: Protection grade
• COUPLING: Serial number of sensor coupled
• ITEM: free for user
The instrument can be installed inside or outside buildings
Altitude: from –200m to 5000m (from -656 to 16400 feet)
Humidity range: 0-100% (IP67/IP65), 0-90% without condensation (IP40) for panel
If the converter is supplied in compact version to the sensor, consider the ambient
temperatures more restrictive, otherwise refer to the respective manuals.
* For discontinuous use, a thermostat heat source installation may be necessary.
** For panel mounting version the maximum operating
temperature is +40°C (104°F)
Instrument classification: class I, IP67 for compact/separate version, IP40 for panel
version (IP65 with optional transparent frontal cover), installation category II, rated
pollution degree 2.
Pmax
44-66Hz 25VA
LV
// 20W
Voltage variations must not exceed ±10% of the nominal one.
All instrument versions can be equipped with all power supply versions.
Input/output are insulated up to 500V
The output 4-20 mA (optional) and the output 24V are electrically connected.