Service manual Leica OH4 Replacing modules/adjusting procedures
Doc. Code 10665744 5-1 04/2007 GOR
5 Replacing modules/adjustment procedures
D-balance module (Mitaka procedure)
1. Turn on the main power and adjust balance correctly by using manual switch.
2. Remove 2 covers, please make switch and switch LED visible (refer to picture 1, 2 & 3).
Picture2
Picture1