EasyManua.ls Logo

Levitronix LEVIFLOW LFC-6CIO - User Manual

Default Icon
31 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
User Manual for LEVIFLOW
®
LFS Flowmeters
www.levitronix.com
PL-4502-00, Rev11, DCO# 18-107
First Release: 29-May-09
Last Update: 30-Jul-18
LEVIFLOW
®
LFS FLOWMETERS
USER MANUAL
LEVIFLOW
®
Sensors: LFS-04(H), -08(H), -20(H), -50(H), -80(H) (LFS-Series)
LFS-008 (Low-Flow Sensor)
LEVIFLOW
®
Converters: LFC-1C (single channel, LFS-Series without LFS-008)
LFC-1C-F4 (single channel, LFS-008 only)
LFC-6C (6-channel, LFS-Series without LFS-008)
LFC-6CIO (6-channel, I/O interface, LFS-Series)
This manual contains information necessary for the safe and proper use of the LEVIFLOW
®
LFS flowmeter series.
Included are specifications for the standard configuration, components, and instructions regarding its use, installation,
operation, adjustment, inspection and maintenance. For special configuration of the LEVIFLOW
®
flowmeters refer to
accompanying information. If the flowmeters have to be configured for other parameter settings then the LEVIFLOW
®
Configuration Software (see relevant manual Levitronix
Doc.# PL-4501-00) is necessary. Please familiarize yourself
with the contents of this manual to ensure the safe and effective use of this product. After reading, please store the
manual where the personnel responsible for operating can readily refer to it at any time.

Table of Contents

Questions and Answers

  • S
    Sharon StoutAug 8, 2025
    Why does the flow indication not match the real flow in Levitronix LEVIFLOW LFC-6CIO?
    • K
      ksmithAug 8, 2025
      If the flow indication on your Levitronix Measuring Instruments doesn't match the actual flow, first, verify that the parameters in the converter are set correctly. Then, inspect the flow path for any accumulation of solids and check the measurement path for excessive bubbles.
  • A
    Angel GalvanAug 12, 2025
    What to do if Levitronix LEVIFLOW LFC-6CIO Measuring Instruments indicate above full scale flow or the flow signal is unstable?
    • D
      Denise SanchezAug 12, 2025
      If your Levitronix Measuring Instruments indicate a flow above full scale or the flow signal is unstable, begin by checking the converter parameters to ensure they are set correctly. Next, inspect the flow path for any solids that may be stagnating and examine the measurement path for excessive bubbles.
  • P
    Phillip RichardsonAug 18, 2025
    What to do if Levitronix Measuring Instruments display a “b-Er” indication?
    • N
      novakpeterAug 18, 2025
      If you see a “b-Er” indication on the display of your Levitronix Measuring Instruments converter, it means the sensor is sending an incorrect signal. Here’s what to check: - Ensure the measuring tube is full of liquid. - Check for obstructions like bubbles or solid substances in the measurement path. - Look for nearby sources of excessive noise, such as motors or high-voltage cables. - Verify that the cables are correctly connected and undamaged.
  • M
    mullinshayleyAug 22, 2025
    Why do Levitronix LEVIFLOW LFC-6CIO Measuring Instruments indicate zero flow when there is flow?
    • D
      David WilliamsAug 23, 2025
      If your Levitronix Measuring Instruments indicate zero flow even when there's flow in the hydraulic circuit, it could be due to incorrect parameter settings, mechanical setup of the sensor, or a zero adjustment performed while there was flow. To resolve this: - Check that the parameters are set to the optimal values. - Ensure the flow direction is correct. - Redo the zeroing process at 0 lpm.
  • P
    patrickayersAug 28, 2025
    How to fix “0-Er” indication on Levitronix Measuring Instruments converter display?
    • A
      Aaron BoydAug 28, 2025
      If the converter display on your Levitronix Measuring Instruments shows a “0-Er” indication, it means the zero adjustment isn't working correctly. Make sure the circuit is filled with liquid and that the flow is actually zero before attempting the zero adjustment again.
  • P
    perezjasonSep 2, 2025
    What to do if Levitronix LEVIFLOW LFC-6CIO indicate zero flow at flow level below cut-off value?
    • M
      millerkevinSep 2, 2025
      If Levitronix Measuring Instruments indicate zero flow at a flow level below the cut-off value, the standard cut-off value might not be suitable for your application. Use the LEVIFLOW® Configuration Software to adapt a lower cut-off value.

Summary

Safety Precautions

General Safety Warnings

Covers hazardous voltage, toxic chemicals, and general caution for safe operation.

Specifications

Installation

LFC-1 C;LFC-1 C-F4 Converter Installation

Covers overview, preparation, electrical, and mechanical installation steps for single-channel converters.

LFC-6 C;LFC-6 CIO Converter Installation

Details installation for 6-channel converters, including electrical and mechanical steps.

Sensor Installation

Guides on sensor overview, preparation, electrical, and mechanical installation for LFS sensors.

Operation

Troubleshooting

Common Troubles

Lists phenomena, causes, and countermeasures for common issues encountered with flowmeters.