LEVIFLOW
®
LFS FLOWMETERS
USER MANUAL
LEVIFLOW
®
Sensors: LFS-04(H), -08(H), -20(H), -50(H), -80(H) (LFS-Series)
LFS-008 (Low-Flow Sensor)
LEVIFLOW
®
Converters: LFC-1C (single channel, LFS-Series without LFS-008)
LFC-1C-F4 (single channel, LFS-008 only)
LFC-6C (6-channel, LFS-Series without LFS-008)
LFC-6CIO (6-channel, I/O interface, LFS-Series)
This manual contains information necessary for the safe and proper use of the LEVIFLOW
®
LFS flowmeter series.
Included are specifications for the standard configuration, components, and instructions regarding its use, installation,
operation, adjustment, inspection and maintenance. For special configuration of the LEVIFLOW
®
flowmeters refer to
accompanying information. If the flowmeters have to be configured for other parameter settings then the LEVIFLOW
®
Configuration Software (see relevant manual Levitronix
Doc.# PL-4501-00) is necessary. Please familiarize yourself
with the contents of this manual to ensure the safe and effective use of this product. After reading, please store the
manual where the personnel responsible for operating can readily refer to it at any time.

Table of Contents

Questions and Answers

  • D
    Dustin LopezAug 8, 2025
    Why does the flow indication not match the real flow in Levitronix LEVIFLOW LFC-1C?
    • T
      Thomas WongAug 8, 2025
      If the flow indication on your Levitronix Measuring Instruments doesn't match the actual flow, first, verify that the parameters in the converter are set correctly. Then, inspect the flow path for any accumulation of solids and check the measurement path for excessive bubbles.
  • P
    Patricia SpencerAug 13, 2025
    What to do if Levitronix Measuring Instruments indicate above full scale flow or the flow signal is unstable?
    • T
      Thomas SmithAug 13, 2025
      If your Levitronix Measuring Instruments indicate a flow above full scale or the flow signal is unstable, begin by checking the converter parameters to ensure they are set correctly. Next, inspect the flow path for any solids that may be stagnating and examine the measurement path for excessive bubbles.
  • R
    Rebecca GardnerAug 17, 2025
    What to do if Levitronix LEVIFLOW LFC-1C display a “b-Er” indication?
    • E
      Emily MorganAug 18, 2025
      If you see a “b-Er” indication on the display of your Levitronix Measuring Instruments converter, it means the sensor is sending an incorrect signal. Here’s what to check: - Ensure the measuring tube is full of liquid. - Check for obstructions like bubbles or solid substances in the measurement path. - Look for nearby sources of excessive noise, such as motors or high-voltage cables. - Verify that the cables are correctly connected and undamaged.
  • K
    Kelsey Henderson MDAug 23, 2025
    Why do Levitronix LEVIFLOW LFC-1C Measuring Instruments indicate zero flow when there is flow?
    • T
      Timothy ReeseAug 23, 2025
      If your Levitronix Measuring Instruments indicate zero flow even when there's flow in the hydraulic circuit, it could be due to incorrect parameter settings, mechanical setup of the sensor, or a zero adjustment performed while there was flow. To resolve this: - Check that the parameters are set to the optimal values. - Ensure the flow direction is correct. - Redo the zeroing process at 0 lpm.
  • S
    skinnerrachelAug 27, 2025
    How to fix “0-Er” indication on Levitronix Measuring Instruments converter display?
    • A
      Ariana SanchezAug 27, 2025
      If the converter display on your Levitronix Measuring Instruments shows a “0-Er” indication, it means the zero adjustment isn't working correctly. Make sure the circuit is filled with liquid and that the flow is actually zero before attempting the zero adjustment again.
  • K
    Kristen GravesSep 1, 2025
    What to do if Levitronix Measuring Instruments indicate zero flow at flow level below cut-off value?
    • C
      carrollchristopherSep 2, 2025
      If Levitronix Measuring Instruments indicate zero flow at a flow level below the cut-off value, the standard cut-off value might not be suitable for your application. Use the LEVIFLOW® Configuration Software to adapt a lower cut-off value.

Summary

Safety Precautions

Specifications

Installation

Installation of Converters LFC-1 C and LFC-1 C-F4

Instructions for installing LFC-1C and LFC-1C-F4 converters.

Electrical Installation for LFC-1 C;LFC-1 C-F4

Step-by-step guide for electrical installation of LFC-1C/LFC-1C-F4.

Mechanical Installation for LFC-1 C;LFC-1 C-F4

Instructions for mechanical installation of LFC-1C/LFC-1C-F4 converters.

Installation of Converters LFC-6 C and LFC-6 CIO

Instructions for installing LFC-6C and LFC-6CIO converters.

Electrical Installation for LFC-6 C;LFC-6 CIO

Instructions for electrical installation of LFC-6C/LFC-6CIO converters.

Mechanical Installation for LFC-6 C;LFC-6 CIO

Instructions for mechanical installation of LFC-6C/LFC-6CIO converters.

Installation of Sensors

Instructions for installing LEVIFLOW® sensors.

Electrical Installation for Sensors

Instructions for electrical installation of LEVIFLOW® sensors.

Operation

Troubleshooting

Levitronix LEVIFLOW LFC-1C Specifications

General IconGeneral
Accuracy±1% of reading
Communication InterfaceRS-485
Power Supply24 VDC
Wetted MaterialsPFA, PTFE
Operating Pressure0 - 7 bar
Output Signal4-20 mA
Flow Range0.2 - 20 L/min
Fluid CompatibilityCompatible with most fluids