Levitronix LEVIFLOW LFSC-i19X - User Manual

Default Icon
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
User Manual for LEVIFLOW
®
LFSC-iX Flowmeters
www.levitronix.com
PL-5001-00, Rev00, DCO# 21-084
First Release: 27-Sep-2019
Last Update: 16-Apr-2021
LEVIFLOW
®
CLAMP-ON FLOWMETERS
USER MANUAL
LFSC-iX Clamp-On Flowmeters
LFSC-i06X:
1/8x1/4”, 1 l/min
1/4x3/8”, 4 l/min
LFSC-i16X:
3/8x5/8”, 20 l/min
LFSC-i19X:
1/2x3/4”, 50 l/min
3/4x1”, 80 l/min
LFSC-i35X:
1x13/8”, 160 l/min
This manual contains information necessary for the safe and proper use of the LEVIFLOW
®
LFSC-iX clamp-on
flowmeter series. Included are specifications for the standard configuration, components, and instructions regarding its
use, installation, operation, adjustment, inspection and maintenance. For special configuration of the LEVIFLOW
®
flowmeters refer to accompanying information. If the flowmeters shall be configured for other parameter settings, then
the Levitronix
®
Service Software (with manual Levitronix
Doc.# PL-4048-00) with version V2.0.6.0 or higher is
necessary. Familiarize yourself with the contents of this manual to ensure the safe and effective use of this product.
After reading, please store the manual where the personnel responsible for operating can readily refer to it at any time.

Questions and Answers

  • J
    jgatesAug 8, 2025
    What to do if Levitronix Measuring Instruments show incorrect flow measurement?
    • E
      Eduardo BarberAug 8, 2025
      If your Levitronix Measuring Instruments display an incorrect flow measurement, it could be due to several reasons: - The measuring tube may not be completely filled with liquid. - There might be disturbances in the measuring path caused by bubbles or solid substances. - Excessive noise from nearby devices like motors or high-voltage cables could be interfering. - Cables might be incorrectly connected or damaged. - The calibration set may not be properly selected. Check these aspects to resolve the issue.
  • H
    hbrownAug 13, 2025
    Why does the flow indication not match the real flow in Levitronix LEVIFLOW LFSC-i19X Measuring Instruments?
    • B
      Blake ReedAug 13, 2025
      The flow indication might not match the real flow in Levitronix Measuring Instruments due to wrong parameter settings or excessive bubbles or solids in the measurement path. To resolve this: - Check if the parameters in the converter are set correctly. - Inspect the flow path for stagnating solids. - Inspect the measurement path for excessive bubbles.
  • C
    Connor SmithAug 18, 2025
    What to do if Levitronix LEVIFLOW LFSC-i19X indicate above full scale flow or the flow signal is unstable?
    • S
      Sandra DuncanAug 18, 2025
      If Levitronix Measuring Instruments indicate a flow above full scale or the flow signal is unstable, it could be due to wrong parameter settings, or excessive bubbles or solids in the measurement path. You should: - Check if the parameters in the converter are set correctly. - Inspect the flow path for stagnating solids. - Inspect the measurement path for excessive bubbles.
  • B
    brandithompsonAug 22, 2025
    How to correct incorrect flow measurement after zeroing Levitronix LEVIFLOW LFSC-i19X?
    • K
      Kevin SanchezAug 23, 2025
      If you're experiencing incorrect flow measurement after zeroing your Levitronix Measuring Instruments, it might be due to a non-ideal situation during the zeroing process. Ensure the circuit is filled with liquid and that the flow is actually zero. Then, perform the zero adjustment again.
  • G
    griffinstevenAug 27, 2025
    What does yellow/red blinking at 1 Hz mean on Levitronix Measuring Instruments?
    • R
      Renee ButlerAug 27, 2025
      A yellow/red blinking light at 1 Hz on Levitronix Measuring Instruments indicates a non-resolvable error, such as an active bootloader or a defect in the primary image. Try power cycling the device, resetting it via the interface, or downloading the correct firmware using Levitronix® Service Software. If the issue persists, the sensor may need replacement.
  • C
    crystal81Sep 2, 2025
    Why is zero flow indicated at a flow level below the cut-off value on Levitronix LEVIFLOW LFSC-i19X?
    • Y
      youngmichaelSep 2, 2025
      If zero flow is indicated at a flow level below the cut-off value on your Levitronix Measuring Instruments, the standard cut-off value may not be adequate for your specific application. Adapt a lower cut-off value with Levitronix® Service Software.

Summary

Safety Precautions

Specifications

Component Overview

Details the parts and their roles in the flowmeter system.

Specifications

Provides detailed technical characteristics of the flowmeter models.

Installation

Preparation

Covers initial steps and checks before installing the flowmeter.

Instructions for Electrical Installation

Provides step-by-step guidance for electrical connections and wiring.

Mechanical and Hydraulic Installation

Details how to physically mount the flowmeter onto the tubing and hydraulic circuit.

Operation

Standalone Operation with User Panel

Explains how to use the flowmeter with its dedicated user panel.

Operation with PLC Interface

Describes how to interface the flowmeter with a PLC system.

Operation with Levitronix Service Software

Covers using software for debugging, data collection, and configuration.

Troubleshooting

Common Troubles

Lists common issues and their causes/solutions.

Troubleshooting with Messages on Flowmeter LED

Explains how to interpret LED status for troubleshooting.

Technical Support

Appendix

Regulatory Status

Details compliance with CE marking and relevant directives.

Disposal of Equipment - WEEE Directive 2012;19;EU

Provides guidance on proper disposal according to WEEE directive.

Symbols and Signal Words

Explains safety symbols and signal word definitions used in the manual.

Levitronix LEVIFLOW LFSC-i19X Specifications

General IconGeneral
BrandLevitronix
ModelLEVIFLOW LFSC-i19X
CategoryMeasuring Instruments
LanguageEnglish