TH 10.211/8.02 - 12/97
11
No.
1
--
-- --
----
--
--
--
--
--
--
2
3
4
5
6
7
9
10
11
12
13
14
8
Activity Response Display on the control panel Fig. 4 Display on the hand unit Fig. 5
Check the mains voltage and
connect the mains cable;
Set the mains switch (3/7) to
the „1“ position
Press the START push-button
(5/3) briefly
--
-- --
--
--
----
Press the STOP push-button
(5/2) for over 1.5 s
Open the test port (3/2)
Disconnect the test sample
Connect the test sample
Press the START push-button
(5/3) briefly
Press the STOP push-button
(5/2) for over 1.5 s
Spray the test sample with
Indication of operational rea-
diness by an acoustic signal.
The first measurement read-
out is obtained at an inlet
pressure for p
E
< 3 mbar
(GROSS mode).)
When the inlet pressure
p
E
< 0.2 mbar the instrument
changes to the FINE mode.
Starting of the stopping pro-
cess.
Starting of the venting pro-
cess.
Display of the leak rate
The first measurement read-
out is obtained at an inlet
pressure for p
E
< 3 mbar
(GROSS mode).)
When the inlet pressure
p
E
< 0.2 mbar the instrument
changes to the FINE mode.
Starting of the measurement
process.
Pumpdown of the dead volu-
me at the inlet.
Initiation of the measurement
process. Pumpdown of the
dead volume at the inlet.
Starting of the stopping pro-
cess.
Starting of the venting pro-
Running up of the forevacu-
um pump and the turbomole-
cular pump. Pumpdown of the
detection system. The auto-
matic self test starts. The
emission is switched on.
„LEYBOLD INFICON“
„UL 200 Version x.x“
Light emitting diode L2 (4/13) is on.
Light emitting diode L3 (4/13) flashes.
„Running Up TMP: 1050 Hz“
„Emission OFF ==> ON“
Light emitting diode L4 (4/13) flashes.
„Standby“
Light emitting diodes L2, L3 and L4
(4/13) are on.
„LR = 3.1E-09 PE = 1.5 E+0“
Light emitting diodes L2, L3, L4 and
L9 are on.
Light emitting diode L9 is off.
Light emitting diodes L2, L3, L4 and
L5 are on.
„Vent“
Light emitting diode L5 is off.
Light emitting diode L7 is on.
Light emitting diodes L2, L3 and
L4 are on.
„Vent“
Light emitting diode L5 is off.
Light emitting diode L7 is on.
Light emitting diodes L2, L3 and L4
are on.
„LR = 5.1 E-07 PE = 1.5 E-1
„LR = 3.1E-09 PE = 1.5 E+0“
Light emitting diodes L2, L3, L4 and
L9 are on.
Light emitting diode L9 is off.
Light emitting diodes L2, L3, L4 and
L5 are on.
Light emitting diode L7 is off.
„Evacuation PE = 3.5 E+1 “
Light emitting diodes L3 and L4 are
on.
Light emitting diode L2 is off.
Light emitting diode L9 is on.
„Evacuation PE = 3.5+1 “
Light emitting diodes L3 and L4
are on.
Light emitting diode L2 is off.
Light emitting diode L9 is on.
All display components are on.
ROM and RAM test.
Decimal point (5/9) flashes.
Test of the display components.
TMP speed is indicated on the
bargraph display (5/4).
Display of helium background or
the detection limit on the bar-
graph display (5/4) and exponent
(5/9).
START display (5/3) is on;
Display of leak rate (5/6) on the
bargraph display (5/4) and expo-
nent (5/9).
--
Display of helium background on
the bargraph display (5/4) and
exponent (5/9);
Vent LED (5/2) is on.
Display of helium background on
the bargraph display (5/4) and
exponent (5/9);
Vent LED (5/2) is on.
START display (5/3) is on;
Display of leak rate (5/6) on the
bargraph display (5/4) and expo-
nent (5/9).
START display (5/3) flashes;
Display of inlet pressure (5/8) on
the bargraph display (5/4).
START display (5/3) flashes;
The pressure in the test sample
(5/8) is indicated on the bar
graph display (5/4).
Blank off the connection flan-
ge (3/2)
Leak rate on bargraph display
(5/4) and exponent (5/9).
2.3 Start-Up (First Pump-Down Cycle)