Installation
17
GA05118_002_C1 - 11/2017 - © Leybold
Operation with additional heat sources involving thermal radiation, thermal
conduction via the high vacuum or the forevacuum flange, strong magnetic
fields or very hot process gases, for example.
Use in systems in which impact stress and vibrations or periodically occur-
ring forces affect pump, frequency converter and cables.
Operation on moving system or system components (locks or movable
pump systems, for example).
Operation at vibration absorbers and vacuum components (gate valves,
valves) which are not capable of sustaining the specified deceleration
torque should the pump rotor seize.
Stepping on pump, add-on parts, drive electronics, flanges and cables to
climb onto the system.
Fitting of add-on parts to the forevacuum flange which cause an inadmissi-
ble high load.
Removing, covering or obstructing warning notices.
Standstill or storing of pump and drive electronics without suitable seal ing-
off and drying. Storing in a humid atmosphere can cause corrosion.
Conversions, manipulations and maintenance work by personnel not
authorised by Leybold.
Any non-conforming utilisation of pump, frequency converter and accesso-
ries can result in severe injury and cause damage to components.