Section 7. Software Reference
Private Network
166.66.77.88
192.168.100.5
Port 7200
192.168.100.6
Port 7200
The public port number should be changed when there is more than one LI-
7200/RS or LI-7500A/RS in the private network. In this case, to connect to each
instrument, two port forwarding rules must be set up, similar to that shown below.
Instrument Public IP Public Port Private IP Private Port
1 166.66.77.88 7201 192.168.100.5 7200
2 166.66.77.88 7202 192.168.100.6 7200
You can run multiple LI-7500RS or LI-7200RS software sessions at the same time
to communicate with different analyzers. Simply double-click on the software icon
to open another session, and connect to a different instrument.
7-6 LI-7200RS Enclosed CO₂/HO Analyzer