Section 7. Software Reference
Baud Rate Maximum Output Rate (Samples per Second)
19200 5
38400 10
57600 15
115200 20
Under Options, you can choose whether or not to output labels with each data
record, and whether to output diagnostic text records. An example of a data record
sent with and without labels is shown below.
Data format with labels:
(Data (Ndx 87665) (DiagVal 757) (Date 2009-09-10) (Time 14:06:44:140) CO2Raw
0.0332911) (H2ORaw 0.19299) (CO2D 5.20672) (H2OD 755.566) (Temp 15.517)
(Pres 99.4361)
Data format without labels:
87665 757 2009-09-10 14:06:44:140 0.0332911 0.19299 5.20672
The End of Line Delimiters determine the character(s) that terminate the data
records. Depending on your recording device, a line feed, carriage return, or both
may be required to properly parse the data records.
Under Data Output, select the data records that you want to output; click Select All
to choose all, or Select None to disable all checked values.
Ethernet Output
The Ethernet tab is used to set the instrument's Ethernet port network com-
munication. After configuration, click Apply and the analyzer will begin to send
data out the Ethernet port according to these parameters.
Use the Update Rate menu to select from .1, .2, .5, 0, 1, 2, 5, 10, or 20 Hz.
Under Options, you can choose whether or not to output labels with each data
record, and whether to output diagnostic text records. An example of a data record
sent with and without labels is shown below.
Data format with labels:
7-14 LI-7200RS Enclosed CO₂/HO Analyzer