If the temperature is 23°C, and the pressure 98kPa, then the mode density is 15.92
mmol CO
2
m
-3
. Then send the following:
(Calibrate(SpanCO2(Target 400)(Tdensity 15.92)(Date "14 Sept 2015")))
The Program Reset Command
(Program(Reset TRUE)) is the equivalent of pressing the reset button on the main
board. It is generally only used to access lower level software for updating the
embedded program, or when the instrument is not responding normally.
Command Remarks
(Program (Reset TRUE)) Sending this will force the software to reset. This is equivalent
to pressing the reset button on the LI-7550 inside panel.
Table D-5. Synchronization Commands
The Network Command
The Name can only include upper and lower case letters (A to Z and a to z), num-
bers (0 to 9), the dash character (-), and the period (.). Names that do not comply
with this constraint will cause communication problems.
Command Subcommands Remarks
(Network (Name {string}) ) Hostname
Address type: dhcp | static
(IP (Type {dhcp | static} )
(Address {xxx.xxx.xxx.xxx})
(Netmask {xxx.xxx.xxx.xxx})
(Gateway {xxx.xxx.xxx.xxx})
(MAC {xx:xx:xx:xx:xx:xx})
Table D-6. Network Commands
The Clock Command
Command Subcommands Remarks
(Clock (Date {}) System date
(Time {}) System time
(PTP {off | automatic | slaveonly | preferred}) Precision Time Protocol setting
(Zone {
1
}) Time zone
1
Refer to the PCsoftware for list of time zones.