General maintenance
If not otherwise stated in the specific product section.
• LINAK
products must be cleaned at regular intervals
• Frequent inspection for malfunction, mechanical damage, wear and cracks. Worn-out parts must be
replaced
• Inspection/maintenance intervals are to be recommended by the medical device manufacturer
• LINAK products are closed units and require no internal maintenance
• LINAK products must be IPX6 Washable and IPX6 Washable DURA when cleaning in wash tunnels
• O-rings: When individual parts are replaced in a LINAK IPX6, IPX6 Washable or IPX6 Washable DURA
system, the O-rings must be replaced at the same time on all parts. On all products where replaceable
cables or fuses have been dismounted or replaced, the O-ring must be replaced, and the O-rings and the
receptacle insert must be greased with an acid-free Vaseline.