21
Setting example 1:
When you want to notify an external measuring instrument that a block check error
has occurred.
Turn on the check mark of the trigger to
be used and touch the trigger factor.
Select "Error" on the trigger factor setting
and check the BCC error you want to detect.
Press [ESC] to return to the trigger
setting screen, then touch the
corresponding trigger action and
select "OT2 pulse output". With this
setting, when a BCC error is detected, it
outputs an L pulse of about 1 ms to the
external trigger output terminal OT2.
Observe the OT2 signal by connecting
measurement device such as an
oscilloscope.
Regardless of the trigger action setting, it outputs outputs an L pulse of about 1 ms to the
external trigger output terminal OT1 when any trigger factor matches.
Setting example 2:
When you want to stop the measurement immediately when receiving the character
string of 41h, 42h, 30h or 41h, 42h, 31h on the transmitting side.
On the setting screen of the checked
trigger factor conguer as shown on the
right. For the third byte of the character
string to be detected we use the bit mask
W0 to set bit 0 to don't care (X).
Press [ESC] to return to the trigger
setting screen and touch the
corresponding trigger action.
Select "Stop measurement" on the
trigger action setting screen.
You can specify whether to stop
measurement immediately or stop after
receiving data by "Stop".