Setting up ECP Water Circulator
Pump
If you have purchased the ECP with your system,
read the following to set up the ECP with the
THMS600 stage. Refer to the ECP manual for
more details.
When heating the stage above 300°C for a pro-
long period of time, the metal casing body of the
stage can get quite hot. Connect the tubing as
shown in the opposite picture.
Note: If you have an LNP95 Liquid Nitr ogen
Cooling system, the tubing from the Dewar must
not be connected to the THMS600 stage when
heating above 300°C. The thin black capillary
tube will melt inside the heater and damage it.
Bypass cooling
ECP Inlet
ECP Outlet