EasyManua.ls Logo

Linkam Scientific Instruments HFS600E-PB4 - Adjusting the Probe on HFS600 E-PB4;PB2; Probe Orientation Method; Adjusting Probe Needle Length; Bending the Probe Needle Tip

Default Icon
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
Adjusting the Probe
Probe orientation:
The probe has a magnetic base (1) and can be
placed in any orientation inside the HFS600E-
PB4/PB2 stage.
The circular metal ring (2) holds the magnetic
foot of the Probe in place.
Adjust the length of the needle:
Pull end of the needle to the desired length and
use a wire cutter to cut the needle shorter.
Bend the tip of the needle:
Use a pair of flat pliers to bend the tip of the
needle.
Note: this action can only be done once. Tung-
sten metal is brittle and bending it again will
break the needle.
1
2

Table of Contents

Related product manuals