BCS196 with Quench Cooling
Option.
Sample Holder Troubleshooting:
It is possible that the retaining catch of the sample
holder may disengage if excessive force is used.
1. Use a set of fine tweezers to grip the
retaining catch.
2. Pull the retaining catch out from the
sample holder.
3. Relocate the retaining catch within
locating slot (also shown on page 16).
Problem - The retaining catch has disengaged
and is resting within the sample holder.
Solution - 1. Grip the retaining
catch with tweezers.
2. Pull the retaining catch out
from the sample holder.
Solution - 3. Relocate the retaining
catch within the locating slot.